5. OPER ATION - ILLUMINATION
t 8IFOVTJOHUIF
Fluorescence Intensity Manager (FIM):
The brightness is adjusted in 5 defined increments (FIM)
using a slider:
100% / 55% / 30% / 17% / 10%
Note:
If the Leica DMi8 is equipped with a 2 port infinity port (
o
S. 44)
switch over to the current port .
o
S. 58
Leica DMi8 with motorized fluorescence
t
If necessary, switch over to the fluorescence axis (Fluor)
with the
TL/Fluor
button.
t 4FUUIFMJHIUJOUFOTJUZPOUIF
Leica EL6000 compact light
source.
(See separate Instructions for Use).
CAUTION
Do not switch on the light source until the light guide is
connected to the microscope.
Uncontrolled light output from the light guide poses a
blinding hazard!
There is also a burn hazard at the light guide output!
Manual integrated fluorescence axis
Manual external fluorescence axis
43