3. SAFE T Y NOTES
3. Safety Notes
3.1
General safety notes
These instruments of protection class 1 are built and in-
spected in accordance with
EN 61010-1 /IEC 61010-1 Safety requirements for
electrical equipment for measurement, control and labora-
tory.
They
also
meet
the
requirements
in
EN
/
IEC
61326-1
Electrical
equipment
for measurement, control and laboratory use – EMC re-
quirements.
The products were also tested in accordance with
EN
62471 / IEC 62471 Photobiological safety of lamps and lamp
systems.
Caution!
In order to maintain this condition and to ensure safe
operation, the user must follow the instructions and
warnings contained in this operating manual.
Caution!
The instruments and accessories described in this In-
structions for Use have been tested for safety and po-
tential hazards.
The responsible Leica Microsystems CMS GmbH af-
filiate or the main plant in Wetzlar must be consulted
whenever the device is altered, modified or used in con-
junction with non-Leica components that are outside of
the scope of this manual.
Unauthorized alterations to the instrument or noncom-
pliant use shall void all rights to any warranty claims!
3.2
Electrical safety
General specifications
Compact Leica CTR electronics box
For indoor use only.
Supply voltage:
100 – 240 VAC
Frequency:
50 / 60 Hz
Power consumption:
max. 150 VA
Fuses:
3.15 A, slow-blowing,
Breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature:
15° - 35°C
Relative humidity:
90% up to 30°C,
non-condensing
Protection class:
I
Overvoltage category:
II
Pollution degree:
2