91
HistoCore PEARL
Cleaning and maintenance
6
Cleaning the retort sieve
Use ethanol or xylene to clean the sieve located at the bottom of the retort. For convenience, the sieve
may be taken out to remove all solid dirt.
Note
Make sure that nothing falls into the drain hole when the sieve is removed for cleaning purpose.
Checking the retort lid seal
Regularly check the retort lid seal for damage. If the seal is damaged, it must be replaced without delay.
The retort lid sealing must be inspected at least once a year. Contact Leica Biosystems Technical
Service for the inspection/replacement service.
Cleaning the display
Clean the screen using a lint-free cloth.
Note
The touchscreen may be triggered during cleaning. Clean the touchscreen whilst there is not a
program currently running.
6.2.3 Monthly cleaning and maintenance
Replacing the active carbon filter
•
The life of the active carbon filter will depend on the reagent types used and the frequency of vacuum
cycles.
•
The filter should be replaced every 45 - 60 days (45 days, warning threshold; 60 days, alarm
threshold). After replacing the filter, open the maintenances menu to reset the filter threshold. For
additional information, see
Note
When inserting the new filter, make sure that the arrows on the filter and on the instrument are in the
same direction. It is inserted with the correct side up, as specified on the label on the front side of the
P. 30 – 4.4.3 Active carbon filter)
Cleaning the reagent bottle
•
Empty and clean the reagent bottles, using a bottle brush in warm water. Wait till they dry and rinse
the reagent bottles with corresponding reagents.
Note
The level can be identified by simply looking through the reagent bottles if these are cleaned
regularly to remove clinging contamination.