5.2.6 System Setup
Setting the paraffin bath temperature
Fig. 41
1
2
3
4
Setting the paraffin bath temperature
L
Only Supervisor and Service user level can edit it.
1. Press the icon of
System Setup
2. Set the paraffin bath temperature, ranging from 50°C - 70°C in 1°C increments
The temperature unit and range is dependant on the settings in
Regional Settings
.
3. Press
Save
Note
It is recommended to set the paraffin baths temperature 2 °C higher than the maximum
paraffin melting point provided in the paraffin manufacturer's specification. Otherwise, the
paraffin that is not completely melted may cause a blockage in the air/liquid system.
If the paraffin baths temperature is set to above 64 °C, the melting time will take longer,
approximately 15 hours.
If the paraffin bath temperature is set from a higher value to a lower value, an error code pops
up. When it occurs, restart the instrument using the On/Off button on the back panel. If any
error appears after the restart, please retry after several minutes. If the problem persists, call
your local service representative.