![Leica BIOSYSTEMS 14 0493 80101 Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/leica-biosystems/14-0493-80101/14-0493-80101_instructions-for-use-manual_1873461062.webp)
62
Version 1.3, Revision D
Operation
5
•
Paraffin baths can only be used for paraffin.
Fig. 48
1
2
3
4
5
Editing the stations
L
Make sure that no program is running.
L
Only Supervisor and Service user level can edit the stations.
1. Press
Stations
.
2. Select a station in the table
3. Press
Edit
, and the reagent list pops up.
4. Select a reagent
. To remove the reagent from a station, select
None
.
5. Press
OK
5.3.4 RMS
The HistoCore PEARL has a Reagent Management System (RMS) that ensures optimized reagent
consumption and the best possible tissue processing results. The Reagent Management System is
controlled via warning threshold values that display the reagent replacement. For each reagent, a value
of this type can be defined depending on the use.
RMS mode
The RMS is divided into 2 groups,
Processing Reagents
and
Cleaning Reagents
.
For
Processing Reagents
, the following 4 modes are available.
•
Off
- The RMS mode of the processing reagents turns off.
•
Cassettes
- Number of processed cassettes
•
Cycles
- Number of cycles (1 processed program = 1 cycle)
•
Days
- Days until changing the reagent
For
Cleaning Reagents
, the following 3 modes are available.
•
Off
- The RMS mode of the cleaning reagents turns off.
Содержание 14 0493 80101
Страница 2: ......
Страница 25: ...25 HistoCore PEARL Initial Operation 4 Fig 3 1 2 3 4 5 6 7 8 9 15 14 ...