31
HistoCore PEARL
Initial Operation
4
Note
The exhausted vapors from the instrument is harmful. The active carbon filter must be in use.
The active carbon filter is only an additional measure to minimize harmful vapors in the area
surrounding the instrument. Ventilation of the workroom is required in any case. The filter
replacement is dependent on the frequency of using the instrument, but at least it needs to be
replaced after 45 - 60 days.
Note
The used carbon filter contains harmful fumes. Dispose of it with proper care according to the
existing applicable local regulations.
4.4.4 External exhaust system
The instrument has been designed so that it can be connected to an external exhaust device. The
optional delivery includes the "Extraction tube"
P. 97 – 8. Ordering information)
for this purpose. It
is recommended to use an external exhaust system.
Fig. 11
1
Attach one end of the extraction tube
to the upwards-pointing tube of the flange
and push it down as far as it will go. Connect the other end of the extraction tube with the external
exhaust station.
Caution
Even if the instrument is connected to an external exhaust device, the active carbon filter provided must
remain in use.
Содержание 14 0493 80101
Страница 2: ......