87
HistoCore PEARL
Cleaning and maintenance
6
6.
Cleaning and maintenance
6.1
Cleaning program
Caution
Do not use the Cleaning program for reprocessing as the Dry step will damage the tissue.
Note
The Cleaning program is a pre-installed program.
The Cleaning program cannot be edited, copied or deleted.
The Cleaning program is offered automatically when an infiltration program has been completed. In
this case, a retort cleaning cycle MUST be carried out; this function cannot be canceled.
Use only the recommended reagents.
Cleaning program (full-step: approximately 1.5 hours))
Step
Reagent
Time (min)
Cycle
Temp. (°C)
Pressure/Vacuum
1
Cleaning xylene
10*
7
67
Ambient
2
Cleaning ethanol
3*
3
62
Ambient
Dry step:
Empty retort
3
Vacuum step
10
80
Vacuum
4
Ventilation step
20
Ambient
Ambient
* Indicating the duration of soaking.
Running Cleaning program
The Cleaning program skips any steps not required to complete the cleaning of the retort.
•
If the last step of a processing program is Formalin, Processing Water or Ethanol, the Cleaning
Program will not run after the processing program completes.
•
If the last step of a processing program is Xylene or Paraffin, the cleaning program will run after the
program completes.
Caution
Do not open the retort lid during the Dry step of the cleaning program.
Caution
Do not pull the reagent bottles when the cleaning program is running. It may cause damage to the
instrument or trigger an error reporting.