ADP300/305 Probe
26
ADP30X-OM-E Rev B
17. Multiply the measured output voltage recorded in step 16 by 100 and divide this number
by the function generator output voltage (probe input voltage) recorded in step 5.
Subtract 1 from this number and multiply the result by 100 to get the error in percent.
100
1
100
%
x
ge
InputVolta
ge
utputVolta
xMeasuredO
Error
⎟⎟
⎠
⎞
⎜⎜
⎝
⎛
−
=
18. Record the answer to two significant places (±x.xx%) on line 18 in the Test Record.
19. Verify that the error is
≤
2.00%.
This completes the Performance Verification Procedure. Complete and file the Test Record, as
required to support your internal calibration procedure. If the criteria in steps 14 or 19 are not met,
contact your local LeCroy service center.