ADP300/305 Probe
16
ADP30X-OM-E Rev B
Probe Offset
PRx:OFFSET, PRx:OFST
Command/Query
Description
When an ADP30X probe is connected to a channel, the OFFSET
command sets the probe offset value.
The maximum range and resolution is determined by the V/DIV setting. If
an out-of-range value is entered, the differential probe will set the offset
to the closest valid value, and the VAB bit (bit 2) in the STB register will
be set.
The OFFSET? query returns the offset voltage of the differential probe
connected to the specified channel.
Command Syntax
<channel>: OFfSeT<offset>
<channel>:={PR1, PR2, PR3, PR4}
ADP305:
<offset>
={-40V to +40V}
for 200 mV/div to 499 mV/div
(LC series) or
={-100V to +100V
} for 200 mV/div to 499 mV/div (LT series) or
={-100V to +100V}
for 500 mV/div to 9.9 V/div or
={-1000V to +1000V}
for 10V/div to 99 V/div or
={-(1400V-4(V/Div setting))
to (1400V-
4(V/Div setting))}
for 100 V/div to 350 V/div
ADP300:
<offset>
={-100V to +100V}
for 1 V/div to 9.9 V/div or
={-1000V to +1000V}
for 10 V/div to 99 V/div or
={-(1400V-4(V/Div setting))
to
(1400V-
4(V/Div
setting))}
for 100 V/div to 350 V/div
Note: The suffix V is optional
Query
Syntax
<channel>:OFfSeT?
Response Format
<channel>:OFST <offset>
Example
The following command sets the offset at the probe tip of the ADP30X
differential probe connected to 1 to 5 volts:
CMD$="PR1:OFST 5": CALL IBWRT(SCOPE%,CMD$)