ADP300/305 Probe
8
ADP30X-OM-E Rev B
SAFETY INFORMATION
To avoid personal injury and to prevent damage to the probe or any products connected to it,
review the following safety precautions. To avoid potential hazards, use the probe only as
specified.
Warning
To avoid fire or personal injury, comply with the following:
For indoor use only.
The probe is intended to be used only with measurement
instruments that are connected to earth ground through the input
BNC connector.
Connect the probe to the measurement instrument before
connecting the test leads to a voltage source.
Do not disconnect the probe from the instrument while the test
leads are connected to a voltage source.
Use of the probe (or the oscilloscope it is connected to) in a
manner other than that specified may impair the protection
mechanisms.
Observe all terminal ratings. Do not apply a potential to any input
that exceeds the maximum rating of that input. See chapter 9 for
detailed specifications of the probe.
To avoid RF burns, do not handle the probe while the input leads
are connected to potentials above the RF Burn Limits shown on
the derating curve (see “Specifications”).
When measuring signals above 2 MHz, be sure to comply with
the voltage vs. frequency derating curve (see “Specifications”).
Otherwise the probe may overheat, thereby compromising the
insulation system.
Avoid exposed circuitry. Do not touch exposed connections and
components when power is on.
Do not use in wet or explosive atmospheres. Remove any
contamination from the probe housing before connecting the