4. Check
Check your CPU manual for the correct trace settings such as port size, port mode and clocking mode.
Wrong settings result in errors.
The trace configuration registers may offer additional help. These are architecture-specific and not always
available. For ETM execute
ETM.Register
:
5. Check
Most of the trace hardware is set up automatically. Special care has to be taken when setting the threshold
level (
Trace configuration
):
The threshold values for clock and/or data signals is/are set automatically at software start, depending on
the voltage level of pin 12 of the target connector. However, if your target is turned off or not connected
during software start, you may need to execute
Trace.THreshold VCC
or manually set it to the proper value
(e.g. 0.9 V , 1.25 V or 1.65 V).
Should be close to 50% of the
ETM signal voltage level.