4.1.3 VACUUM CHUCK WET TEST
To prevent chemicals from entering into the vacuum path, the chuck and o-ring
surface must be clean and defect free. A chuck wet test can be performed to
check the integrity of the o-ring seal. See Fig.4-2. Using a clear substrate, place
the wafer on the chuck and press the vacuum key. Using a bottle of DI water
spray water around the periphery of the substrate where it meets the chuck while
manually rotating the chuck. Check to see if there is any leakage across the o-
ring.
If there is no leakage, the chuck is ok to use.