Proline Kryomats
64
Starting up
YAUE0007 / 10/2/2018
7.6.4
Restoring works settings
All works settings, apart from the control parameters and the sensor calibration, are restored.
Command
Works settings
Open the device parameter menu via
the soft key Menu .
With the cursor keys continue to
Settings
Works settings
.
The window shown opposite appears.
Master and then
only control par.
int.
is shown as a possible choice.
There are however various possibilities,
which can be selected with
or
:
Under
All modules
Master, Command
and all connected modules are reset to
the works setting with
all default
.
Under
Master
you have the choice
between:
all default
, then all Master settings
are reset,
only control para. int.
for the
internal control parameters,
only control para. ext.
similar for
external,
only miscellaneous
which resets
setpoint, pump level, maximum
current consumption, control to
internal and auto start to "Auto".
Under
Command
all command settings
are reset with
All default
.
Confirm selection with
.
Confirm the control dialog shown on the
left with
or cancel with
.
Return to measurement window with
End or
.
Pump Menu
End
T
set
T
fix
All modules
Master
Command
Cool
all default
only control par. int
only control par. ext
only miscellaneous
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel
Содержание RP 3050 C
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Страница 2: ......
Страница 4: ......
Страница 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Страница 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Страница 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Страница 134: ......
Страница 135: ......