Proline Kryomats
YAUE0007 / 10/2/2018
Starting up
83
7.11.1
Internal control variable (integral measurement probe)
Only read further here, if you have no external temperature probe connected (and activated according
to Section 7.5.4 as control variable).
Command
Control Parameters
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Control
Control Parameters
Control Parameters
.
The adjacent window appears.
Change parameters marked with (auto)
where necessary to manual input with
Tv manual/auto
.
Select the parameters to be changed with
and confirm with
.
Then in the following settings window,
change the value and confirm with
.
7.11.1.1 Proven settings for control parameters and pump (integral measurement probe)
Instrument
Type
Heat transfer liquid
Xp
_P
Tn
tn
Tv
tu
Td
td
Pump level
RP 4090 CW
Water
9.0
60
42
6.3
8
RP 4090 CW
Ethanol
9.0
60
42
6.3
8
RP 4090 C
Water
9.0
60
42
6.3
8
RP 4090 C
Ethanol
9.0
60
42
6.3
8
Technical changes reserved!
7.11.2
External control variable (External measurement probe)
You only need to read further here if you have connected an external temperature probe or the actual
temperature is read in from a module (and you have activated it as control variable according to Section
7.5.4).
Only modify the control parameters if you have knowledge of control techniques.
The control system for external actual values is implemented for improvement of the control behavior as
a two-stage cascade controller. A "master controller" determines the "internal setpoint", from the
temperature setpoint and the external temperature, passed to the slave controller. The control value of
the slave controller controls the heating and cooling.
When a setpoint step change is specified, it may be that the optimum control would set a bath
temperature, which might significantly exceed the temperature desired on the external vessel. There is
a correction limitation, which specifies the maximum permissible deviation between the temperature on
the external load and the heat transfer liquid temperature.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Xp
6,0
Tn
30
Tv
(auto)
21
Td (auto)
3,5
Содержание RP 3050 C
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Страница 2: ......
Страница 4: ......
Страница 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Страница 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Страница 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Страница 134: ......
Страница 135: ......