Proline Kryomats
96
Starting up
YAUE0007 / 10/2/2018
Warnings in the
“Command-Display”
Warnings from
“Cooling system”
Message
Meaning
Message
Meaning
WX201
Overflow of CAN receipt
WX301
Overflow of CAN receipt
WX202
Watchdog-Reset
WX302
Watchdog-Reset
WX203
RTC Voltage drop recognized: Battery failure
WX303
sm.stell_min still not determined
Adaption run
necessary
WX304
Pressure switch 1 operated
WX305
Condenser dirty (
cleaning)
WX210
Software version of control system too old
WX310
Software version of control system too old
WX211
Software version of protection system too old
WX311
Software version of protection system too old
WX213
Software version of heating system too old
WX312
Software version of operation system
WX214
Software version of analogue interface too old
WX314
Software version of analogue interface too old
WX215
Software version of RS-232 too old
WX315
Software version of RS-232 too old
WX216
Software version of contact I/0 too old
WX316
Software version of contact I/0 too old
WX217
Software version of valve 0 too old
WX317
Software version of valve 0 too old
WX218
Software version of valve 1 too old
WX318
Software version of valve 1 too old
WX219
Software version of valve 2 too old
WX319
Software version of valve 2 too old
WX220
Software version of valve 3 too old
WX320
Software version of valve 3 too old
WX221
Software version of pump 0 too old
WX321
Software version of pump 0 too old
WX222
Software version of pump 1 too old
WX322
Software version of pump 1 too old
WX223
Software version of pump 2 too old
WX323
Software version of pump 2 too old
WX224
Software version of pump 3 too old
WX324
Software version of pump 3 too old
Warnings from “Analogue-Module”
Warnings from
“RS232/485-Module”
Message
Meaning
Message
Meaning
WX401
Overflow of CAN receipt
WX501
Overflow of CAN receipt
WX402
Watchdog-Reset
WX502
Watchdog-Reset
WX410
Software version of control system too old
WX510
Software version of control system too old
WX411
Software version of protection system too old
WX511
Software version of protection system too old
WX412
Software version of operation system
WX512
Software version of operation system
WX413
Software version of heating system too old
WX513
Software version of heating system too old
WX415
Software version of RS232 too old
WX514
Software version of analogue interface too old
WX416
Software version of contact I/0 too old
WX516
Software version of contact I/0 too old
WX417
Software version of valve 0 too old
WX517
Software version of valve 0 too old
WX418
Software version of valve 1 too old
WX518
Software version of valve 1 too old
WX419
Software version of valve 2 too old
WX519
Software version of valve 2 too old
WX420
Software version of valve 3 too old
WX520
Software version of valve 3 too old
WX421
Software version of pump 0 too old
WX521
Software version of pump 0 too old
WX422
Software version of pump 1 too old
WX522
Software version of pump 1 too old
WX423
Software version of pump 2 too old
WX523
Software version of pump 2 too old
WX424
Software version of pump 3 too old
WX524
Software version of pump 3 too old
Содержание RP 3050 C
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Страница 2: ......
Страница 4: ......
Страница 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Страница 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Страница 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Страница 134: ......
Страница 135: ......