Preparation
39
The external application must not have a hydraulically blocking effect.
If cross-sectional area of tube is too low
temperature gradient between bath
and external load due to low flow rate.
Always ensure the largest possible passages in the external circuit.
When tightening the union nuts on the pump nipple AF 19, use a wrench AF 14
to counter the tightening torque (see figure).
If external control is to be used, provide a Pt100 temperature probe in the
external load (
With loads at a higher position and with stationary pump and ingress of air into
the thermostatic circuit, the external volume can drain away, even with closed
circuits
Risk of thermostat overflowing!
Secure hoses against slippage with hose clips.
Unused pump connectors must be closed off.