Proline Kryomats
YAUE0007 / 10/2/2018
Maintenance
119
Master: Alarm message
PulEU
Command: Low level (pump)
(Pump motor monitoring: No load).
1. No liquid in the system. If this occurs, the level
monitoring has failed.
Check whether the float
in the expansion vessel is blocked by foreign
bodies. Otherwise, contact LAUDA Service
Constant Temperature Equipment
2. With the option "open load" the device draws air
out of the open load.
Move the return to the
load.
Master: Alarm message
Error 11
Command: Overpressure
(outflow pressure too high).
Pump level too high
Select a lower pump level
Compressor overtemperature
Three-phase current
9.4
Disposal information
The following applies for EU member states: The device must be disposed of
according to Directive 2012/19/EU (WEEE Waste of Electrical and Electronic
Equipment).
9.4.1
Disposal of the refrigerant
. The type and filling quantity of the refrigerant can be read on the unit or on the rating plate. Repair
and disposal only through a qualified refrigeration engineer!
The following applies for EU member states: The disposal of the refrigerant must be carried out
according to EC Directive 2015/2067/EU in conjunction with 517/2014/EU.
9.4.2
Disposal of the packaging
The following applies for EU member states: The disposal of the packaging must be carried out
according to the EC Directive 94/62/EC.
Содержание RP 3050 C
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Страница 2: ......
Страница 4: ......
Страница 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Страница 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Страница 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Страница 134: ......
Страница 135: ......