Proline Kryomats
YAUE0007 / 10/2/2018
Maintenance
113
9.2.2
Draining the water-cooled condenser
Important:
With the risk of frost (e.g. transport in winter), drain the condenser on
water-cooled devices.
Remove the water hose on the water tap. Than open the solenoid valve for the water as described
below. Blow compressed air in the water return hose Continue until all water has flowed out of the
device.
Command
Start unfill condenser
Open the device parameter menu via the
soft key Menu .
Change from
Pump
Start unfill conden.
using
.
Use
to open the solenoid valve for
the water. The display changes
automatically to the overview window.
To close the valve again selects this
menu again. The item is called now
End unfill conden.
.
9.3
Servicing, cleaning, repair and disposal information
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Pump Menu
End
T
set
T
fix
Pump Level
Ext.Pump
Start unfill conden
Pump
Содержание RP 3050 C
Страница 1: ...2SHUDWLQJ LQVWUXFWLRQV 352 1 U RPDWV RZ WHPSHUDWXUH 7KHUPRVWDWV 53 53 53 53 53 53 53 53 ...
Страница 2: ......
Страница 4: ......
Страница 40: ...Proline Kryomats 40 Starting up YAUE0007 10 2 2018 7 Starting up ...
Страница 41: ...Proline Kryomats YAUE0007 10 2 2018 Starting up 41 ...
Страница 132: ...Proline Kryomats 132 YAUE0007 10 2 2018 ...
Страница 134: ......
Страница 135: ......