Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
95
Command
Explanation
IN_PAR_06
Read current value of TvE (0 = OFF).
IN_PAR_07
Read current value of TdE.
IN_PAR_09
Read current value of correction limitation
IN_PAR_10
Read current value of XpF.
IN_PAR_14
Read setpoint offset.
IN_PAR_15
Read current value of PropE.
IN_DI_01
State of contact input 1: 0 = open / 1 = closed.
IN_DI_02
State of contact input 2: 0 = open / 1 = closed.
IN_DI_03
State of contact input 3: 0 = open / 1 = closed.
IN_DO_01
State of Contact output 1:
0 = make-contact open / 1 = make-contact closed.
IN_DO_02
State of Contact output 2:
0 = make-contact open / 1 = make-contact closed.
IN_DO_03
State of Contact output 3:
0 = make-contact open / 1 = make-contact closed.
IN_MODE_00
Keyboard Master: 0 = free / 1 = inhibited.
IN_MODE_01
Control: 0 = internal / 1 = external Pt100 / 2 = external analogue /
3 = external serial
IN_MODE_02
Standby: 0 = Unit ON / 1 = Unit OFF.
IN_MODE_03
Keyboard Command: 0 = free / 1 = inhibited.
IN_MODE_04
Setpoint offset source: 0 = normal / 1=external Pt / 2 = external analogue /
3 = external serial
TYPE
Read the device type
(for example “RP 845”).
VERSION_R
Read software version number of control system.
VERSION_S
Read software version number of protection system.
VERSION_B
Read software version number of Command.
VERSION_T
Read software version number of cooling system.
VERSION_A
Read software version number of analogue module.
VERSION_V
Read software version number of RS-232/485 module.
VERSION_D
Read software version number of digital module.
VERSION_M_0
Read software version number of solenoid valve (Cooling water).
VERSION_M_1
Read software version number of solenoid valve (Automatic refill).
VERSION_M_3
Read software version number of solenoid valve (shut-off valve 1).
VERSION_M_4
Read software version number of solenoid valve (shut-off valve 2).
VERSION_M_5
Read software version number of high-temperature cooler.
STATUS
Read equipment status 0 = OK, -1 = error.
STAT
Read error diagnosis
Response: XXXXXXX
X = 0 no error, X = 1 error.
1. char = error
2. char = Alarm
3. char = Warning
4. char = over temperature
5. char = low-level error
6. char = high-level error (at adjustment alarm)
7. char = no external control variable