Proline Edition X cooling thermostats
76
YACE0100 / October 4, 2018
3.
Setting the master controller (PIDT
1
-controller):
Start with setting Auto and proceed with Manual only when necessary.
3.1. Kpe setting:
Start with the data from table 7.13.2.1.
In case of oscillations with large period, i.e.10 min)
Kpe higher, otherwise lower.
3.2.
– Tne/ Tve/ Tde setting:
Start with the data from table 7.13.2.1; and with high numbers
(Tne = 70 s
— 200 s; Tve = 50 s — 150 s).
With lower numbers
faster approach, otherwise slower approach with lower oscillations.
Tve: to reduce overshot
Tve higher, otherwise lower.
Tde (damping for Tve): in general approximately 10 % of Tve.
4.
Correction limitation (or outflow temperature limitation) (
7.13.2.1) und temperature limits (Til/Tih)
(
Make settings in accordance with the boundary conditions. Examples:
Heat transfer liquid
Correction limitation
Til
Tih
Water
depending on the external
vessel size and the
thermostatic liquid
2 °C
95 °C
Ethanol
Minimum
40 °C
Tools to watch the time behavior:
Graph mode of the Command remote control,
LAUDA Wintherm PC-program.