Command
T
set
or T
fix
or the soft key T
set
opens the
setpoint window.
123.45
is the setpoint, which is still active.
The upper and lower limit temperatures
are displayed (device-specific values).
There are three different possible entry
methods:
1. Change the value with the
or
keys. First, you vary the 1/10 °C values. If
you hold the key pressed longer, then full
degrees change.
2. Enter the complete number with the
numerical duo keys and the
key for the
negative sign and decimal point.
3. Using
or
, move the flashing cursor
line to the decimal place which you would
like to change and then change it with
or
.
Confirm the value with
or quit the
window with
without having made any
changes.
Two other ways of entering the setpoint:
With the soft key T
fix
open the
window shown on the left.
The setpoints, which you last entered, are
shown in the right-hand column. In the
illustrated screen, the last setpoint was
80.0 °C.
To accept an earlier setpoint, enter the
right-hand column with
and select the
desired value with
, then accept it with
or cancel with
.
In the left-hand column setpoint
temperatures, which are to be used
frequently, can be defined as "fixed
settings".
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Enter new setpoint:
123,
45
Min: -40.00°C Max:202.00°C
Fixed settings
Recent setpoints
0.00°C
80.00°C
0.00°C
-35.50°C
0.00°C
20.00°C
0.00°C
38.00°C
0.00°C
-35.70°C
0.00°C
0.00°C
0.00°C
0.00°C
0.00°C
0.00°C
Pump Menu