Proline Edition X cooling thermostats
74
YACE0100 / October 4, 2018
Parameter
Sl
ave controller (P-controller):
Sl
proportional range:
iP
= Xpf
in Kelvin.
Tve, Tde, Prop_E changeover to Auto/ Man
E
A
= automatic or
E
MN
= manual.
Works setting is automatic. These three parameters should only be
modified by experienced control technicians!
Hold
pressed for 3 seconds, then make changes.
Correction limitation
Et0l
See introduction (
Command
Control
Control Parameters
Control
Parameters
Open the device parameter menu via the
soft key
Menu
.
With the cursor keys continue to
Control
Control Parameters
Control Parameters
.
The adjacent window appears.
Parameter endings:
e = Master controller, f = Slave controller.
Where applicable change parameters
marked with (auto) to manual input with
Tv manual/auto
.
Select the parameters to be changed with
and confirm with
.
Then change the value in the following
settings window and confirm with
.
Correction limitation
see introduction
(
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Correction limitation
Kpe
0,50
Tne
100
Tve (auto) 83
Tde (auto)
8,3
Xpf
4,0
Prop_E(a) 30
Et0l
°C
E A
°C