Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
55
Confirm the control dialog on the right
with
or cancel with
.
Return to the measurement window with
End or
.
7.8.9 Entering the offset of the external temperature probe
If a deviation is found during the check using a calibrated reference thermometer probe, then the offset
(the additive part of the characteristic) of the external measurement chain can be adjusted with the
following function. The reference thermometer must be dipped nearly by the external temperature probe
into the consumer bath according to the details on the calibration certificate.
Master
CAL E
3x
to
MNEnu
2x
to
CAL..
2x
Call the offset adjustment for the external temperature probe
CAL E
.
Continue as described in (
7.8.7) for the internal temperature
probe.
Command
Calibration
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Settings
Calibration
extern Pt100
Calibration
.
The adjacent window appears.
Confirm selection with
.
Continue as described in (
7.8.7) for the
internal temperature probe.
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default