Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
49
Command
Offset source
and
Setpoint offset
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Control
Setpoint offset
Offset
source
.
Off
indicates that the setpoint offset is
currently deactivated.
Select the setpoint source with
or
and confirm with
.
Interfaces (e.g. RS-232) are only
displayed if a valid setpoint has already
been transmitted.
With the cursor keys continue to
Setpoint offset
.
The standard value is
0.00°C
.
Open the left-hand window with
.
Enter the desired temperature.
Accept the change with
,
quit the window with
without making
changes.
Pump Menu
End
T
set
T
fix
Off
extern Pt100
RS232
Offset source
Setpoint offset
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Input Setpoint offset:
0.
00
Min: -500.00°C Max: 500.00°C
Pump Menu
End
T
set
T
fix
0.00 °C
Offset source
Setpoint offset
Содержание Proline RP 1290 C
Страница 2: ......
Страница 4: ......
Страница 115: ...Proline Edition X cooling thermostats YACE0100 October 4 2018 115 Blank page ...
Страница 116: ...Proline Edition X cooling thermostats 116 YACE0100 October 4 2018 Blank page ...
Страница 118: ......
Страница 119: ......