SET-HD100
20
(E)
8-2. Installation Error Range and Environmental and Rotational
Accuracy Conditions for Guaranteeing Accuracy
Item
Specifi cation
Notes
Circular run-out
of scale relative to
rotational axis
When using a mechanical
reference to check the
scale unit inner diameter
part
0.010 mm
Install using the reference surface of
the scale unit inner diameter part as
the reference, or use a microscope
and install using the lines recorded
on the scale surface as the reference.
See section 3-1-3 “Example of
Installation Procedure” for details
of the lines recorded on the scale
surface.
When using a microscope
and checking with the
line references recorded
on the scale surface
0.020 mm
Total run-out of scale mounting surface
0.010 mm
Including installation tolerance
and fl uctuation during rotational
movement.
Circular run-out of head unit mounting surface
0.050 mm
Including installation tolerance
and fl uctuation during rotational
movement. See section 9
“Dimensions” for details.
Co-axiality of head
unit relative to
scale unit
Head unit inner diameter
reference
0.040 mm
Installation using the supplied axis
alignment jig is possible. See section
9 “Dimensions” for the reference
surface and other details.
Head unit outer diameter
reference
0.020 mm
Installation using an external
reference is possible. See section
9 “Dimensions” for the reference
surface and other details.
Distance between scale and head reference
surfaces
1.503 ±0.020 mm
Including installation tolerance
and fl uctuation during rotational
movement.
Axial run-out of stage
Less than 0.010 mm
Rotational accuracy during self-calibration
1 to 10 min
-1
Calibration may not be possible if the
rotational accuracy during
self-calibration is low.
Tightening torque
during installation
Head unit (M6)
1.0 ±0.1 N·m
Control with a torque wrench.
Uneven or excessive tightening
torque may reduce the accuracy or
cause damage.
Scale unit (M4)
0.6 ±0.06 N·m
8-3. Calibration Conditions
Th
e calibration conditions of the issued calibration certifi cate are shown below.
Item
Condition
Notes
Temperature
20 ±1°C
Number of angle points
1,048,576
Measured signals
A, B and Z signals
Th
e rising edge of the A signal is set as
the angle point. Angle point No. 1 is set
at the rising edge of the fi rst A signal aft er
detection of the Z signal (reference signal).
Rotation conditions
CW, CCW
5 min
-1
Mounting position
Standard mounting position
described in section 9 “Dimensions”
Angle deviation and uncertainty are
calculated through measurement of a total
of 16 installation status patterns consisting
of four diff erent scale unit and head unit
positions each shift ed by 90° symmetric to
the rotational axis.
Содержание SET-HD100
Страница 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Страница 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Страница 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Страница 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Страница 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Страница 36: ...26 J SET HD100 ...
Страница 60: ...SET HD100 16 E Carry case Bands Lock Secure with the bands 6 Close and lock the lid Storage complete 7 ...
Страница 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Страница 70: ...SET HD100 26 E ...
Страница 71: ......