SET-HD100
8
(E)
Installation
1
Temporarily mount the scale unit
Set the scale unit on the mounting surface of the rotating part, check the installation position with the
supplied axis alignment jig, and then semi-tighten the four screws.
*
Use the screws that were attached to the scale unit case, or screws procured by the customer.
Scale unit
Rotating part
Screw
Screw
2
Adjust the eccentricity of the scale unit.
Rotate the rotating part and adjust the eccentricity of the temporarily mounted scale unit while
checking the eccentricity accuracy. Adjust by such as pressing lightly on the side of the scale unit base,
taking care not to damage the machine.
<Specifi cation>
See section 8 “Specifi cations.”
<Examples of check methods>
• Use a dial test indicator or other tool and measure the circular run-out of the reference surface
(inner diameter part) of the scale unit base.
• Use a microscope and measure the concentric circles on the scale scanning surface.
Scale unit
Press lightly on the side
of the scale unit base.
Dial test indicator
Microscope
(approximately ×100 to ×1000)
Rotating part
Rotational axis of rotating part
R40.223
Concentric circle pattern within
scale increments
Line width: 2 μm
R40.8
R42.4
R42.6
R42.8
R41.0
R41.2
R41.4
R41.6
R41.8
R42.0
R42.2
R40.6
R48.223
Содержание SET-HD100
Страница 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Страница 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Страница 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Страница 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Страница 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Страница 36: ...26 J SET HD100 ...
Страница 60: ...SET HD100 16 E Carry case Bands Lock Secure with the bands 6 Close and lock the lid Storage complete 7 ...
Страница 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Страница 70: ...SET HD100 26 E ...
Страница 71: ......