7. Trouble Information
Check the following items and take the proper remedial actions when the status indicators of the
interpolator unit indicate an error or an alarm.
Symptom
Cause
Remedial action
Th
e STATUS indicator lights
up or blinks red.
Self-calibration failed.
See section 3-1 “Installation” and
section 4-1 “Measurement when
Performing Self-calibration,” and
perform installation again.
A SENSOR (1 to 6) indicator
lights up red.
A level alarm or speed alarm has
occurred.
In case of a level alarm:
Th
e sensor signal level has dropped.
a
Dropped due to dirtiness
See section 6-1 “Cleaning” and clean
the scale unit.
b
Th
e installation position may not
satisfy the tolerance range.
See section 3-1 “Installation” and
section 8-2 “Installation Error
Range and Environmental and
Rotational Accuracy Conditions for
Guaranteeing Accuracy,” and check
the installation position.
In case of a speed alarm:
Th
e maximum rotary response speed
was exceeded.
Use at the maximum rotary response
speed or less.
Th
e POWER indicator is off .
Th
e power is not turned on.
Check that power supply (+24 V) is
input. Check whether the fuse has
melted.
If performing the above actions does not resolve the trouble, contact our service department.
Содержание SET-HD100
Страница 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Страница 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Страница 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Страница 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Страница 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Страница 36: ...26 J SET HD100 ...
Страница 60: ...SET HD100 16 E Carry case Bands Lock Secure with the bands 6 Close and lock the lid Storage complete 7 ...
Страница 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Страница 70: ...SET HD100 26 E ...
Страница 71: ......