- 37 / 79 -
LANGER
EMV-Technik
DE-01728 Bannewitz
[email protected]
www.langer-emv.de
FLS 106 IC set / FLS 106 PCB set
The height-adjustable machine feet allow the user to adjust the scanner to uneven work surfaces.
Figure 35: Machine feet on the scanner
The feet can be set to the desired height with the height adjustment screw (hexagon socket 4 mm).
The clamping screw (hexagon socket 2.5 mm) fixes the feet at the respective height.
The adjustable height is 3.8 mm.
Instructions on how to change the height of the feet:
- Loosen the height-adjustment lock by turning the clamping screw counter-clockwise. Turn the
height-adjustment screw (hexagon socket 4.0 mm) clockwise to raise the scanner. Turn the
height-adjustment screw counter-clockwise to lower the scanner. Tighten the clamping screw
after adjusting the correct height.
Give the scanner enough time to acclimatize before putting it into service if it is brought from a cold
into a warm environment. Condensation may settle down on the scanner's surface. The
condensation should be promptly removed with dry, clean cloths. Acclimatization may take up to
several hours depending on the temperature difference. We recommend that further steps only be
taken after the scanner has adjusted to the working area's ambient temperature.
Langer EMV-Technik GmbH assumes no liability for damage to property and/or personal injury
or for consequential damage that arises from the incorrect unpacking or installation of
the FLS 106 scanner.