SR1 Operation
97
© 2014 Stanford Research Systems
Configuring Sweep Data
After a new sweep X-axis point is determined according to the selected Sweep Source, SR1 begins to
search for settled values of each of the sweep data measurements. Up to 6 measurements may be
selected for each sweep. To select the sweep data measurements, click on the numbered box in the
sweep data section of the sweep panel. This opens a tree display of all the available measurement
selections. Click on the measurement to add it to the sweep, or click on "None" to remove the current
measurement from the sweep. Only measurements from the currently active analyzers will be shown, so
it's important to set up the
analyzers
before configuring the sweep.
Other Sweep Parameters
"Pre-Sweep Delay" sets a fixed delay between the time when the sweep starts and the time when the
analyzer begins measruing the first data point. "Measurement Timeout" sets the maximum time that the
analyzer waits for settled measurements, if a settled measurement is not returned in this time interval
the analyzer skips that point and moves on. Checking "Rpt" (Repeat) causes the sweep to automatically
restart as soon as it finishes the current sweep.
Sweeps and Events
SR1's
Event
panel allows a number of user-defined actions to be associated with the occurrence of
certain events. Actions include logging the occurrence to a text file, running a script, or triggering a COM
event. Sweeps define several such events: Sweep Start, Sweep New Point Start, Sweep New Point
Timeout, Sweep New Point Done, and Sweep Finished.
Содержание SR1
Страница 5: ...Part I Getting Started Audio...
Страница 7: ...Getting Started 7 2014 Stanford Research Systems...
Страница 12: ...SR1 Operation Manual 12 2014 Stanford Research Systems...
Страница 27: ...Part II SR1 Operation Audio...
Страница 156: ...SR1 Operation Manual 156 2014 Stanford Research Systems Passband Group Delay of Elliptical Filter...
Страница 258: ...SR1 Operation Manual 258 2014 Stanford Research Systems...
Страница 272: ...SR1 Operation Manual 272 2014 Stanford Research Systems on the amplitude sweep...
Страница 289: ...SR1 Operation 289 2014 Stanford Research Systems...
Страница 290: ...Part III SR1 Reference Audio...