SR1 Operation
257
© 2014 Stanford Research Systems
2.8.3
Reference Panel
The
Reference
Quick Measurement uses the
Time Domain Detector
to measure the output of the EUT
at a specified amplitude and frequency and record the result in the dBr references for each output
channel.
Cabling Configuration
The output of SR1's analog or digital generator is connected to the EUT's input. The output of the EUT is
connected to SR1's analog or digital inputs. The number of inputs and outputs and the connectors used
are specified in the
Setup Panel
.
Generator Configuration
The output waveform is set to either Sine or Low Distortion Sine as specified in the panel.The amplitude
and frequency of the generator is set to the frequency and level.
Analyzer Configuration
One or both analyzers (depending on the number of output channels specified in the
Setup Panel
) are
configured as
Time Domain Detectors
. The bandwidth limits for the measurement is selected on the
"Analysis" section of the panel.
Free Run Configuration
A barchart display is created for each output channel showing the instantaneous amplitude for the two
inputs.There is no sweep configuration for the Reference measurement.
Содержание SR1
Страница 5: ...Part I Getting Started Audio...
Страница 7: ...Getting Started 7 2014 Stanford Research Systems...
Страница 12: ...SR1 Operation Manual 12 2014 Stanford Research Systems...
Страница 27: ...Part II SR1 Operation Audio...
Страница 156: ...SR1 Operation Manual 156 2014 Stanford Research Systems Passband Group Delay of Elliptical Filter...
Страница 258: ...SR1 Operation Manual 258 2014 Stanford Research Systems...
Страница 272: ...SR1 Operation Manual 272 2014 Stanford Research Systems on the amplitude sweep...
Страница 289: ...SR1 Operation 289 2014 Stanford Research Systems...
Страница 290: ...Part III SR1 Reference Audio...