SR1 Operation
283
© 2014 Stanford Research Systems
2.9.3
Digital-Analog
The Digital-Analog selection lists basic Audio Analyzer measurements with a digital stimulus and an
analog response, i.e. a DAC. The
digital generator
is configured with the stimulus signal. The sampling
rate of the high-resolution analog converter is set to the digital OSR, so both domains are operating with
the same samplling rate.
What you'll need to configure for all these setups:
Change the digital output parameters (connector, sample rate, bit resolution) to match the input of the
DACt being tested. Set the
analog inputs
to the appropriate connector and input coupling.
FFT
This setup configures SR1 for a basic audio band FFT. The Hi-resolution converter is selected for
maximum dynamic range. The frequency range of the FFT is 1/2 the sampling rate of the digital output.
The A0 analyzer is configured as a dual channel FFT analyze continuous averaging.. The Digital
Generator is loaded with a lsine waveform set to a default frequency of 1 kHz and an amplitude of 1 FFS.
Page2 contains a graph displaying the A and B channel power spectra while Page 3 contains a graph
displaying the A and B channel time records.
What you'll need to configure:
If your'e using the digital generator, adjust the amplitude and frequency of the signal to match your
requirements. Otherwise you can turn the generator off. The amount of
averaging
in the FFT analyzer can
be adjusted to suit the noise level of your signal. The FFT resolution is defaulted to 4k points, adjust this
to obtain the proper balance of update rate vs. frequency resolution
Frequency Response
This setup configures SR1 for a stereo audio-band frequency response sweep. The Hi-resolution
converter is selected for maximum input dynamic range. The Time Domain Detector is selected for the
A0 analyzer. The Digital Generator is loaded with a sine waveform. Page2 contains a graph displaying
the A and B channel frequency response. The sweep controller is programmed to sweep the sine
frequency from 20 Hz to 20 kHz in 22 logarithmic steps.
What you'll need to configure:
Adjust the the sine amplitude to match the equipment being tested. The number of points in the sweep
can be changed to give faster sweeps or higher resolution sweeps. The
settling parameters
can be
adjusted to fit the noise levels of the signals being measured.
Wideband FFT
This setup is similar to the audio-band FFT setup described above except that the analog input sampling
rate is set to OSRx2 so that the FFT bandwidth is twice as large. The A0 analyzer is configured as a
dual channel FFT analyzer with continuous averaging.. The Digital Generator is loaded with a low-
distortion sine waveform set to a default frequency of 1 kHz and an amplitude of 1 Vrms. Page2 contains
a graph displaying the A and B channel power spectra while Page 3 contains a graph displaying the A
and B channel time records.
What you'll need to configure:
If your'e using the digital generator, adjust the amplitude and frequency of the signal to match your
requirements. Otherwise you can turn the generator off. The amount of
averaging
in the FFT analyzer can
Содержание SR1
Страница 5: ...Part I Getting Started Audio...
Страница 7: ...Getting Started 7 2014 Stanford Research Systems...
Страница 12: ...SR1 Operation Manual 12 2014 Stanford Research Systems...
Страница 27: ...Part II SR1 Operation Audio...
Страница 156: ...SR1 Operation Manual 156 2014 Stanford Research Systems Passband Group Delay of Elliptical Filter...
Страница 258: ...SR1 Operation Manual 258 2014 Stanford Research Systems...
Страница 272: ...SR1 Operation Manual 272 2014 Stanford Research Systems on the amplitude sweep...
Страница 289: ...SR1 Operation 289 2014 Stanford Research Systems...
Страница 290: ...Part III SR1 Reference Audio...