SR1 Reference
301
© 2014 Stanford Research Systems
Analog Signal Meters
RMS Level Meter
Accuracy (1 kHz ref)
±0.5 % (±0.043 dB)
Flatness (rel. 1 kHz,
Amp. = 4 Vrms)
20 Hz to 20 kHz <±0.008 dB (typ. < ±0.003 dB)
10 Hz to 64 kHz <±0.020 dB
10 Hz to 200
kHz
<±0.030 dB
Frequency Meter
Range
8 Hz to 300 kHz
Accuracy
±(2 ppm + 10 mHz)
Phase Meter
±1.0° accuracy
Digital Signal Meters
Frequency Meter
10 Hz to 0.45 F
s
, ±100 ppm accuracy
Phase Meter
±0.5° accuracy (f = 50 Hz)
Analyzers (Analog and Digital Audio)
Time Domain Analyzer
Measurements
Amplitude, amplitude ratio, THD+N, THD+N ratio, SINAD
Analog Inputs:
Amplitude Accuracy
±0.5 % (±0.043 dB)
Flatness (1 kHz ref)
50 Hz to 20 kHz
<±0.008 dB (typ.±0.003 dB)
20 Hz to 64 kHz
<±0.02 dB
10 Hz to 200 kHz
<±0.05 dB
Residual Noise
(62.5 mVrms Input Range, Shorted Input)
High Res. ADC (Fs = 128
kHz)
22 Hz to 22 kHz
–117.5 dBu
22 kHz to 57.6 kHz
–115 dBu
A-Weighted
–120 dBu
High BW ADC
22 Hz to 22 kHz
–118 dBu
22 kHz to 80 kHz
–113 dBu
10 Hz to 200 kHz
–110 dBu
A-Weighted
–120 dBu
Residual THD+N
High Res. ADC
(Fs = 128 kHz)
Содержание SR1
Страница 5: ...Part I Getting Started Audio...
Страница 7: ...Getting Started 7 2014 Stanford Research Systems...
Страница 12: ...SR1 Operation Manual 12 2014 Stanford Research Systems...
Страница 27: ...Part II SR1 Operation Audio...
Страница 156: ...SR1 Operation Manual 156 2014 Stanford Research Systems Passband Group Delay of Elliptical Filter...
Страница 258: ...SR1 Operation Manual 258 2014 Stanford Research Systems...
Страница 272: ...SR1 Operation Manual 272 2014 Stanford Research Systems on the amplitude sweep...
Страница 289: ...SR1 Operation 289 2014 Stanford Research Systems...
Страница 290: ...Part III SR1 Reference Audio...