Table B-1
MPS Mainframe Control Bus Remote Command Summary
Input
Command
Query Function
*CLS
–
Clear all pending status.
*ESE
*ESE?
Event status register enable.
– *ESR?
Event
status
register contents.
– *IDN?
Product
identification.
*SRE
*SRE?
Service request (status byte) enable.
– *STB?
Status
byte
register contents.
–
*TST?
Self-test error status.
MODE
MODE?
IEEE interface mode status (not used).
TERM
TERM?
IEEE interface terminator status. Terminators fixed at (CR)(LF).
I, ISET
I?, ISET?
Output current setting.
V, VSET
V?, VSET? Output compliance voltage setting.
–
IOUT?
Output current measurement.
–
VOUT?
Output voltage measurement.
–
?
Output current, voltage, status byte, and programming mode.
IMAX
IMAX?
Maximum output current setting limit.
RAMP1
RAMP?
Ramp segment parameter programming.
SEG
SEG?
Active ramp segment (fixed at 1).
RMP
RMP?
Ramp active enable and status.
VCAL
ICAL
CALPL
CALMN
VZER
IZER
CALZ
–
Mainframe digital calibration.
ZI
ZI?
Output current zero setting.
ZIS
ZIS?
Output current zero enable and status.
–
IMODE?
Current programming mode switch status.
–
VMODE?
Voltage programming mode switch status.
– RES?
Normal or high resolution status.
–
OVP?
Output overvoltage protection status.
– ERR?
Error
Status.
–
RI?
Remote inhibit status.
ISTP
ISTP?
Output current step setting.
ISTPS
ISTPS?
Output current step enable and status.
STEPR1
–