Lake Shore Model 647 Magnet Power Supply User’s Manual
Remote Operation
4-15
4.3.1.3 Ramping Commands Description
The following MPS Ramping Commands configure the various functions of the ramping features.
RAMP:
Programs Ramp Segment Parameters
Input:
RAMP[segment],[initial ramp current],[final ramp current],[ramp rate],[00],[--:--:--:--]
where:
[segment]
=
1
. (Future updates will allow up to nine segments.)
[initial ramp current]
=
0 to ±72 A
.
[final ramp current]
=
0 to ±72 A
.
[ramp rate]
= 0 to 99.9999 A/S.
[00]:
Reserved for future use (Operation to perform). Need not be present for
proper command operation.
[--:--:--:--]:
Reserved for future use (DWELL time in days, hours, minutes and
seconds). Does not need to be present for proper command operation.
Remarks:
Truncates value to 0.001 place. Enter values with no embedded spaces. Undefined
parameters are set to 0. Control Bus inputs 4 places, display rounds to 0.01 or 0.001 place.
Example:
RAMP1,+72.0000,-72.0000,01.0000[term]
Indicates:
Ramp Segment = 1;
Initial Ramp Current = +72.0000 A;
Final Ramp Current = -72.0000 A;
Ramp Rate = 01.0000 A/S; commas can be replaced by blanks.
RAMP?:
Ramp Parameter Query
Input: RAMP?
Returned:
Values are the same format as described above for programming in the ramp segments.
Remarks:
Returns 48 characters and up to two terminators.
RMP:
Programs Ramp Status
Input:
RMP [ramp status]
where
[ramp status]
=
0
to turn off (HOLD) the ramp definitely, or
1
to
turn on the ramp segment or continue a ramp on hold.
RMP?:
Ramp Status Query
Input:
RMP?
Returned:
0
for holding, or
1
for ramping.
Remarks:
One character plus up to two terminators are returned.
SEG:
Programs Active Ramp Segment
Input: SEG
[segment]
where
[segment]
=
1
. (Future updates will allow multiple ramp segments).
SEG?:
Active Ramp Segment Query
Input: SEG?
Returned:
1
. (Future updates will allow multiple ramp segments).
Remarks:
Returns one character plus up to two terminator.