59
Communications Interface
Instructions for Setup
RS232 and RS485 serial communications are available. They are accessible via the DB-9 connectors
on the rear of the chiller. Refer
Inadmissible Operating Conditions
A Terminal Emulator or other comparable device will need to be connected to the DB-9 to allow
command to be entered.
Terminal Settings:
Baud Rate 115200, Data – 8-bit, Parity – none, Stop – 1 bit, Flow Control - none
Commands and responses:
Command
Description
General Response
CMP
Compressor on//off
Status of Compressor query (returns "ON/OFF")
COL
Set/Get Coolant type
Coolant Percentage (returns "xx")
CRT
Coolant Return Temp
Coolant Return Temperature query (returns “xxx.x C/F”)
CSP
Coolant supply pressure
Coolant Supply Pressure query (returns “xx.xx bar/psi”)
CST
Coolant Supply Temp
Coolant Supply Temperature query (returns “xxx.xx C/F”)
CTL
Set/get control loop
System Control Loop [R {Run}, S {Stop}]
DAT
Set/Get date
Set/Get Date (returns "xx/xx/xx)
FCC
Fan duty cycle
Fan Duty cycle query (returns xxx.xx)
FLO
Set/Get flow option
Flow sensor Availability check (returns "1" if present, "0" if not present)
FLW
Get flow rate
Flow Rate query (returns “xx.xx LPM/GPM”)
LVL
Coolant level sensor
Coolant Level Switch status query (returns “LOW”, “OK”)
MOD
Model
Displays the model number
PBP
Pressure bar/psi
Set Pressure in bar/psi (returns "bar/psi")
PMP
Pump on//off
Control Power to Pump [ON, OFF] (default = OFF)
RDP
Get R_CSP
Refrigerant Discharge Pressure query (returns “xx.xx bar/psi”)
REF
Refrigerant type
Refrigerant type R513A
RSP
Get R_SSP
Refrigerant Suction Pressure query (returns “xx.xx bar/psi”)
SCF
Degrees C or F
Select °C or °F [C, F] (default = C)
SSN
System serial number
Serial number is displayed
TIM
Set/Get time
Time [00:00:00-23:59:59] (default = 00:00:00)
TSP
Temp setpoint
Control Temperature Set Point [-10.0 to 40.0] (default = 20)
VER
Get firmware version
Version query (returns “FIRMWARE VERSION X.X”)
XPM
Flow Rate Unit
Flow Rate Unit [G {Gallons}, L {Liters}] (default = L)
Command entry format
$[2-character source ID] [3 Character Command], [Data Field if required]
Command strings without contents in Data Field will be interpreted as a query.
Source ID
PC
Serial link to a PC or external digital serial interface
Command Structure Examples
$PCCST
(Requests current Temperature of Fluid Supply Line via PC)
CST,25
(Response from controller reporting Coolant Supply Temperature)
$PCVER
(Returns [ VER FIRMWARE VERSION x.x.x])
Содержание Nextreme NRC1200 Series
Страница 1: ...Nextreme Recirculating Chiller Specification and User Manual Part Number 387004848 ...
Страница 17: ...17 Component Locations Figure 2 Isometric Views of Unit Figure 3 Top View ...
Страница 18: ...18 Figure 4 Front View Figure 5 Rear View Figure 6 Side View ...
Страница 37: ...37 Figure 21 Refrigerant Circuit and Coolant Circuit Figure 22 Coolant Circuit and Control Panel ...
Страница 39: ...39 Figure 25 Sensor Wiring Low Voltage ...
Страница 40: ...40 Plumbing Refrigeration Diagram Figure 26 Plumbing and Refrigeration Diagram ...
Страница 43: ...43 Figure 31 Pump Performance of NRC5000 with Cooling Fluids Water and 40 EG W ...
Страница 60: ...60 PCFLT Returns FLT fault or NO FAULTS FOUND PCFLT C Clears Fault Returns FLT Faults Cleared ...
Страница 69: ...69 Declaration of Conformity ...