16
TECHNICAL SPECIFICATIONS
Model
NRC
50
00
Performance
Cooling Capacity
1
5,000 Watts
(17,060 BTU/hr)
Setpoint Range
-10°C to 40°C
(14°F to 104°F)
Temperature Stability
±0.1°C (±0.2°F)
Nominal Flow Rate
1
(50Hz / 60Hz)
15 lpm @ 1.7 bar / 15 lpm @ 2.8 bar
(3.9 gpm @ 25 psi / 3.9 gpm @ 41 psi)
Maximum available pressure
5.3 bar (77 psi)
Refrigerant
R 513A
Storage
Temperature, w/o coolant
-25°C to 70°C
(-13°F to 158°F)
Humidity
5% to 95%, non-condensing
Operation
Coolant
Water or Water/Glycol
Temperature
2
15°C to 40°C
(59°F to 104°F)
Relative Humidity
30% to 80%
Altitude
≤2,000 meters
≤(6,560 feet)
Input
Voltage
220-230 VAC
Frequency
50/60 Hz
Physical
Dimensions, W x D x H
63 x 59 x 91 cm
(24.8 x 23.2 x 35.8 in)
Weight (w/o coolant)
100 kg (220 lbs)
Coolant Capacity
5 L (1.3 gal)
Couplings
1/2” NPT
Compliance
•
UL Mark for Laboratory Equipment
(ANSI / UL / CSA / IEC EN 61010-1 Edition
3)
1. Nominal capacity rating is given at a 20°C (68°F) setpoint, 20°C (68°F) ambient temperature, sea level, 60Hz operation,
and the ST1 pump.
2. For ambient conditions outside this range, please contact Laird Thermal Systems.
Содержание Nextreme NRC1200 Series
Страница 1: ...Nextreme Recirculating Chiller Specification and User Manual Part Number 387004848 ...
Страница 17: ...17 Component Locations Figure 2 Isometric Views of Unit Figure 3 Top View ...
Страница 18: ...18 Figure 4 Front View Figure 5 Rear View Figure 6 Side View ...
Страница 37: ...37 Figure 21 Refrigerant Circuit and Coolant Circuit Figure 22 Coolant Circuit and Control Panel ...
Страница 39: ...39 Figure 25 Sensor Wiring Low Voltage ...
Страница 40: ...40 Plumbing Refrigeration Diagram Figure 26 Plumbing and Refrigeration Diagram ...
Страница 43: ...43 Figure 31 Pump Performance of NRC5000 with Cooling Fluids Water and 40 EG W ...
Страница 60: ...60 PCFLT Returns FLT fault or NO FAULTS FOUND PCFLT C Clears Fault Returns FLT Faults Cleared ...
Страница 69: ...69 Declaration of Conformity ...