Preparing the Unit for Maintenance
All necessary safety measures must be taken to prevent accidents when carrying out the maintenance.
The following preparations must be made:
Terminate the cooling operation.
Let the unit and its coolant cool down to the ambient temperature.
Switch off the unit.
Disconnect the unit from mains by pulling the mains plug.
Secure the unit against being switched on again.
Verify that the unit is de-energized.
Keep unauthorized persons away from the working area.
Place the unit on a level surface.
Verification of Safe State after Maintenance
Visually inspect the refrigerant line to make sure there are no leaks.
Ensure that all screws are securely fastened
Make sure that there are no loose electrical connections.
Ensure that the controller completes the diagnostic.