
58
Alarm
Criticality
Alarm Description
Cause
Effect
Action Required/ Troubleshooting
Sel
f-T
es
t
Self-Test
Red
Controller checks
whether all the sensors
are functioning properly
after the power switch is
turned on
Sensors not
connected properly
or component
failure
System
wouldn’t Run if
self-test fails
Check if there are any other alarms
listed in the alarms page of the LCD
screen along with this alarm. Check
if those issues can be resolved
using the guidelines above. After
fixing the issue, acknowledge the
alarm on the LCD screen. This will
clear the alarm. Now Run the unit. If
the problem persists, contact LTS
customer service.
C
ool
ant
T
em
per
at
ur
e
Coolant
temperature
alarm
(customer)
Red
Customer method to
detect high temperature
Coolant return
(inlet) temperature
is higher than 50°C
Depends on
the method
customer
implements
Verify the heat load has not
exceeded the rating of the
recirculating chiller.
If the unit is still powered, check the
alarm screen to find the referenced
alarm in this table.
After fixing the issue, acknowledge
the alarm on the LCD screen. This
will clear the alarm. Now run the
unit. If the problem persists, contact
LTS customer service.
If the unit was shut down, repower
and run the unit. If the problem
persists, contact LTS customer
service.
Содержание Nextreme NRC1200 Series
Страница 1: ...Nextreme Recirculating Chiller Specification and User Manual Part Number 387004848 ...
Страница 17: ...17 Component Locations Figure 2 Isometric Views of Unit Figure 3 Top View ...
Страница 18: ...18 Figure 4 Front View Figure 5 Rear View Figure 6 Side View ...
Страница 37: ...37 Figure 21 Refrigerant Circuit and Coolant Circuit Figure 22 Coolant Circuit and Control Panel ...
Страница 39: ...39 Figure 25 Sensor Wiring Low Voltage ...
Страница 40: ...40 Plumbing Refrigeration Diagram Figure 26 Plumbing and Refrigeration Diagram ...
Страница 43: ...43 Figure 31 Pump Performance of NRC5000 with Cooling Fluids Water and 40 EG W ...
Страница 60: ...60 PCFLT Returns FLT fault or NO FAULTS FOUND PCFLT C Clears Fault Returns FLT Faults Cleared ...
Страница 69: ...69 Declaration of Conformity ...