Indications for Use
The BAG resuscitator is a self-inflating, manual resuscitator that is intended for patients requiring total or
intermittent ventilation support.The BAG resuscitator provides positive pressure ventilation and also allows
for spontaneous breathing either with an artificial airway or with a face mask (3)(a), (b) or (c).
Cautions and Warnings
• Resuscitators should only be used by persons who have received specific training in their use.
• When using supplemental oxygen, do not allow smoking or use unit near sparking equipment, open
flame, oil or other flammable chemicals.
• The BAG should not be used in toxic or hazardous atmospheres.
• Infant and Child units are equipped with a pressure-limiting device (2) which opens at a pressure of
approx. 35±5 cmH2O. However, an abrupt, high-volume manual breath may cause the unit to exceed this
level.The pressure relief valve can be overridden by depressing the plunger momentarily with light finger
pressure or locked in place by using the push-twist locking feature.
• The resuscitator is for single patient use only. Do not reuse. Do not sterilize.
• Use the appropriate resuscitator size (8)(a), (b) or (c).The wrong size can result in inadequate or
excessive air pressure being delivered to the patient. See the Performance and Specifications table for
sizing.
• When using supplemental oxygen, the flow from its source should be monitored. Delivery of
supplemental oxygen greater than 30 LPM (Liters Per Minute) may result in inadvertent positive
end expiratory pressure (PEEP).
• The use of third party products and oxygen delivery devices (e.g. filters and demand valves) with The
BAG Disposable Resuscitator may have an affect on product performance. Please consult with the
manufacturer of the third party device to verify compatibility with The BAG and obtain information on
the possible performance changes.
Limited Warranty
The BAG Disposable Resuscitator is warranted against defects in workmanship and materials only.
Please refer to the Global Warranty statement for additional terms and conditions (www.laerdal.com).
Practical Operation
1. Remove the resuscitator from the outer protective poly bag. Expand the Adult or Child resuscitator
from collapsed configuration to its operating position.
2. Inspect the unit to be sure the system is complete.
3.
PRE-OPERATIVE FUNCTIONALTEST:
• Compress the ventilation bag (8) with one hand then release the grip on the bag. Rapid bag
re-expansion confirms efficient air intake.
• Block the patient valve/mask connector part and try to compress the bag. If the bag cannot be
compressed with reasonable force, the valve is efficiently preventing backward escape of air.
• Place a reservoir bag, (6) or (7), or test lung (if available) over the patient valve. Compress the
bag several times.This should fill the reservoir or test lung and confirm that the patient valve is
able to efficiently direct air to the patient.
• Compress the filled reservoir bag.Air should vent to the atmosphere as indicated by lifting of the
disk membrane at the base of the mask connector and not return to the ventilation bag (8).
4.
PRESSURE RELIEF REGULATOR
:The Infant and Child resuscitators feature a patient valve with a
special pressure limiting device (2) mounted on the upper valve housing. If inspiration meets with
pulmonary resistance of approximately 35±5 cmH2O, the device opens,reducing the risk of stomach
distension and / or barotrauma. A hissing sound can be heard when the device opens. If higher
ventilation pressure is required, the pressure limiting device can be overridden with finger pressure on
the plunger or disabled by depressing and turning the plunger.
5. If high concentrations of oxygen are needed, attach oxygen tubing (4) to the bag and an adjustable
oxygen source, and attach an oxygen reservoir bag (6) or (7) to the bag.
6. Adjust oxygen flow to insure the oxygen reservoir bag remains fully or partially inflated during use.
7. When using a mask (3) attached to the resuscitator, be sure you have a tight seal and secure fit.
8. When using an endotracheal tube or tracheotomy tube,remove mask (3) and attach the resuscitation unit
directly to the tube.The mask elbow provides a 15 mm I.D.port for this purpose.
9.
HAND STRAP
: Grab the bag with your hand.Adjust the hand strap (9) by loosening the loop from the
hook then pull the strap to fit your hand, and finally reattach the loop to the hook.
The BAG II rev C 01:The BAG II rev C 30.11.2007 09:29 Side 4