APPENDIX 1. MENUS AND TOOLBARS
Menu Functions
The
File menu
lets you open, close and save sequence files (.seq files). You can also choose the Trace Measurements option,
which lets you record voltage and current for each channel, over time, to an external file.
The
Edit Menu
lets you cut, copy and paste text as you edit your sequence.
The
View menu
lets you set up the Sequence interface:
-
Toolbar
displays or hides Sequence’s tool bars.
-
Status
displays or hides the status bar at the bottom of the Sequence window.
-
Split
lets you resize the two panes of the Sequence window.
-
Options
>
Communications
lets you set up communications with the HVS448.
-
Options
>
Monitoring
lets you choose options for monitoring HVS functions.
-
Options>HVS448 Model
lets you pick the HVS448 Model number for the unit you are controlling.
-
Options>Show All Dialogs
turns on all dialog boxes so they will all appear.
-
Options>Active Voltage Control
is the default that shows current and voltage of all channels and lets you change
current and voltage by dragging.
The
Actions menu
controls communication with the HVS448 as well as the progress of the current sequence.
-
Enable High Voltage
applies high voltage to the channels.
-
Disable High Voltage
stops high voltage from being applied even when a voltage is set for a channel.
-
Program
All
sends the current settings and sequence to the HVS448.
-
Run
lets you start the current sequence at any of its steps, A–H.
-
Pause
halts progress of the current sequence. All channel states remain as they are. This is equivalent to the
Pause/Resume
button on the HVS448 front panel.
-
Resume
restarts progress of the current sequence. This is equivalent to the Pause/Resume button on the HVS448 front
panel when the sequence is paused.
-
Reset
/
Stop
ends the current sequence and sets all channels to inputs.
-
Update
Firmware
lets you update the firmware in the HVS448 as new features become available.
-
Get
Versions
reports the current versions of firmware and software.
-
Calibration>Recalibrate
performs a series of internal calculations to calibrate the voltage settings. While the HVS is
performing this function, the high voltage channels are supplying up to the maximum voltage to the back-panel high
voltage connectors.
Be sure to disconnect everything from the high voltage connectors on the back panel before
performing a calibration
. Following recalibration, the channels are inputs until they receive new settings.
-
Calibration >Save Calibration
saves the calibration results to the HVS.
-
Calibration> Restore Calibration
resets the HVS448 to a saved calibration.
Lock Front Panel
prevents accidental or purposeful adjustment of sequences or voltages from the front panel. When the panel is
locked, the LED next to the power button glows a steady red and all buttons are dark.
Unlock Front Panel
allows front panel control of the HVS448. This is the default condition. When the panel is unlocked, the
power LED glows green and the front panel buttons are illuminated.
The Lock functions are nonvolatile: the panel will remain locked or unlocked even if power is removed and restored. The panel
can only be unlocked from within Sequence software.