17
-
Sequencer
opens the
Sequencer Control Panel
, an overview of the current sequence’s settings
-
Sequencer A
–
H
opens the sequence for an individual channel
-
Step A
–
H
shows a step within a sequence
-
Programmable Trigger Logic
lets you define the logic of the signal sent to the digital input.
Figure 10
. Sequence Main Window, showing navigation in the left pane and channel settings on the right
Monitor Settings
Monitoring channel measurements and status will help you interpret the state of the HVS channels. Choose
View>Options>Monitoring
to control how voltage, current and sequence status will be monitored (
Figure 11
):
Figure 11. Monitor Settings
-
Measurement interval
is the frequency with which measurement values are updated in milliseconds. The measurement
interval must be an integer in the range of 1 to 1000000000.The new measurement interval will be applied the next time
Sequence starts. To apply the new measurement interval immediately, select the new interval, click
OK
, set the software to
monitor
None
(see
Manual
Power Supply Control Panel) then reselect the channels to monitor.
-
Sampling period
is the default period over which measurements are integrated in the HVS. The default period is 25.6 ms;
you can also select a shorter sampling time from the drop-down list.
Note
: During high speed measurements the added processing required for monitoring measurements may add jitter to
your timing. Use monitoring thoughtfully if you are performing high-speed or timing sensitive experiments.