Level
Designation / function
Settings / descriptions
Process input A and B can be either a sensor 1 or a sensor 2. Further information about the type of sensor 1 or 2 please
refer to the signal converter manual "Sensor input combinations". Process input A is always present, process Input B
can be present.
Note: The exchange of a sensors 1 with a sensor 2, or vice versa, can only be done by the manufacturer!
Depending on the sensor which is connected to a slot A or B the menu changes.
C1.1
C1.1
C1.1
C1.1
parameter (cond. ind.)
This menu item is for selecting the probe which is connected to process
input A/B. The entries of this selection depends on the chosen device
configuration. The device configuration is customer specific and set during
production.
C1.2
C1.2
C1.2
C1.2
cell constant
Enter cell constant.
C1.14
C1.14
C1.14
C1.14
time constant
Enter time constant.
C1.15
C1.15
C1.15
C1.15
temperature
Menu for temperature measurement. Available for sensor 1 and sensor 2.
C1.15.1
probe
Options:
•
manual: used if no internal or external temperature sensor is connected to
the signal converter
•
Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter
C1.15.2
manual
Only available if C1.15.1 is set to "manual".
C1.15.3
correction
Offset correction for temperature measurement. Not available if C1.15.1 is
set to "manual". If you have chosen
“
Pt 1000
”
enter the temperature
correction.
C1.15.4
limitation
Measuring ranges for temperature measurement.
C1.15.5
temp. comp.
Menu for activating the temperature compensation parameters for the
measurement.
Options:
•
Linear: linear temperature compensation.
•
Off: temperature compensation is disabled.
C1.15.6
temp. coefficient
If you have chosen
“
linear
”
enter the temperature coefficient.
C1.15.7
ref. temperature
If you have chosen
“
linear
”
enter the reference temperature.
C1.23
C1.23
C1.23
C1.23
cell calibration
For detailed information refer to Calibration refer to
Calibrating
measurement
on page 26
C1.23.1
temp. comp.
Menu for temperature measurement.
Options:
•
off: temperature measurement is disabled.
•
manual: temperature value has to be entered manually.
•
automatic: temperature measurement is performed as configured
C1.23.2
temperature
Menu for setup of manual temperature (manual temperature
measurement).
C1.23.3
temp. coefficient
Menu for setup of temperature coefficient (manual temperature
measurement).
C1.23.4
spec. conductivity
Enter the conductivity of the calibration fluid.
Menu for preparation of measurement:
•
Put probe into calibration fluid
•
Wait until measurement is stable.
•
Press enter to proceed.
Wait until concentration is measured.
C1.23.5
cell constant
Enter cell constant.
Check slope, press enter, decide whether to store or to discard calibration
parameter.