OPERATION
4
25
OPTISENS IND 1000
www.krohne.com
12/2016 - 4001969202 - MA OPTISENS IND 1000 R02 en
4.3.2 Calibrating the temperature measurement
This section is only required if the measured temperature is not correct.
The temperature is commonly measured with the integrated temperature probe and therefore
step 2a does apply in most cases. In rare cases of a process being very stable in temperature
manual compensation is used and step 2b applies.
Step 2: correcting the temperature measurement
•
If necessary, enter the temperature correction in Kelvin so that the signal converter shows the
same temperature as the reference thermometer. Press
^
to confirm the entered value. The
temperature sensor has been adjusted.
Step 2a: probe Pt1000
probe Pt1000
probe Pt1000
probe Pt1000
Read the currently measured temperature of the Pt1000 temperature sensor from the measurement
screen and write it down.
Measure the temperature with a reference thermometer and check if it deviates from the temperature
measured by the Pt1000.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu setup
setup
setup
setup is highlighted.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,
beneath the submenu process input A
process input A
process input A
process input A is highlighted. Choose process input A or B where ever
pH is configurated.
•
Press
or
to select process input A
process input A
process input A
process input A.
•
Press >>>> to enter the chosen menu .
You are on the second submenu level. Press
or
until the submenu temperature
temperature
temperature
temperature is
highlighted.
•
Press >>>> to enter the chosen menu.
You are on the third submenu level. Press
or
until the submenu correction
correction
correction
correction is
highlighted.
•
Press >>>> to enter the chosen menu.
•
Press
or
and
to change the reading.
•
Press
^
to confirm the value