Enter the measured temperature. Press
^
to confirm the entered value. The manually
measured temperature will now be used for temperature compensation.
4.3.3 Calibrating measurement
Verification and in case of need calibration is necessary in regular intervals or good practice
when installing a new sensor. A sensor should be recalibrated if the verification within a
standard solution shows a too high difference in actual reading.
When calibrating a conductivity measurement, keep in mind that the measuring system as a
whole is calibrated, and not only the electrode.
To avoid alarms on the distrubted control system (DLC) when temporarily removing the sensor
(i.e. for maintenance), the converter has a hold function. This function "freezes" all outputs (i.e.
the display and the current outputs) of the last measured value.
Step 2b: probe manual
probe manual
probe manual
probe manual
Measure the temperature of the measuring medium.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu setup
setup
setup
setup is highlighted.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,
beneath the submenu process input A
process input A
process input A
process input A is highlighted.
•
Press
or
to select process input A
process input A
process input A
process input A.
•
Press >>>> to enter the chosen menu process input A
process input A
process input A
process input A.
You are on the second submenu level. Press
or
until the submenu temperature
temperature
temperature
temperature is
highlighted.
•
Press >>>> to enter the chosen menu. The submenu probe
probe
probe
probe is highlighted.
•
Press
or
until the submenu manual
manual
manual
manual is highlighted.
•
Press
^
to enter the chosen menu
INFORMATION!
As an indication that the manual hold function is active, the "warning sign" in the upper left
corner of the display appears. Meanwhile the status messages show "checks in progress". For
more details about how to select the manual hold function refer to the signal converter manual.