OPERATION
4
27
OPTISENS IND 1000
www.krohne.com
12/2016 - 4001969202 - MA OPTISENS IND 1000 R02 en
After starting-up the signal converter, the measuring screen appears. This is the standard
screen which is displayed automatically in the normal operating mode. If you are in this mode
and you want to initiate a calibration, you have to activate the manual hold function in the first
step.
Step 1: activating the manual hold function
•
You have activated the manual hold function. To go to the next step and prepare the calibration
procedure. You have to return to the measuring mode.
•
Press
^
until you reach the measuring mode again
Step 2: preparing the calibration procedure
•
If you re-calibrate an existing sensor, remove the sensor from its respective assembly
(for further information refer to
Mounting to a flow through assembly
on page 17 or refer to
Calibrating measurement
on page 26).
•
If you calibrate a sensor, make sure that the sensor is correctly connected to the converter.
•
Check the sensor for damages or dirt deposits.
•
During the calibration procedure, you will have to dip the sensor into a standard solution.
Provide a suitable vessel.
After activating the manual hold function and the preparative measures, you can get access to
the calibration procedure from the measuring mode via the main menu setup
setup
setup
setup (step 3).
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
•
Press
or
until the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
•
Press >
>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "quick setup" and
A1" appears, beneath the submenu language
language
language
language is highlighted.
•
Press
or
until the submenu manual hold
manual hold
manual hold
manual hold is highlighted.
•
Press >
>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display
"manual hold" appears, beneath the option off
off
off
off is highlighted
•
Press
or
to choose the option on
•
Press
^
to confirm the entered value.