Temp Probe Adjustment
Display
Action
Remark
Select Calibration.
Press
enter
to proceed.
Select CAL_RTD
calibration method.
Press
enter
to proceed.
Wrong settings
change the
measurement
properties!
Measure the tempera-
ture of the process
medium using an
external thermometer.
Display (3 sec)
Now the device is in
HOLD mode.
Enter the measured
temperature value.
Maximum difference:
10 K.
Press
enter
to proceed.
Display of actual
temperature (un-
compensated) in the
lower display.
The corrected tempera-
ture value is displayed.
Sensoface is active.
To end calibration:
Select MEAS,
then
enter
To repeat calibration:
Select REPEAT,
then
enter
After end of calibra-
tion, the outputs re-
main in HOLD mode
for a short time.
After calibration is
ended, the device will
switch to measuring
mode.