Calibration by Entry of an Installation Factor
When using a Memosens sensor in a tight space, the installation factor
is entered.
Display
Action
Remark
Select Calibration.
Press
enter
to proceed.
Select CAL_INSTALL
calibration method.
Press
enter
to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Enter installation factor.
Press
enter
to proceed.
The selected pro-
cess variable and
the temperature are
displayed.
Use the arrow keys to
select:
• MEAS (end)
• REPEAT
Press
enter
to proceed.
End:
HOLD is deactivated
after a short time.