59
Configuration
Menu item
Action
Choices
TAG
Select ON or OFF using
keys.
Press
enter
to confirm.
When switched on, the
entry for “TAG” in the
Memosens sensor is com‑
pared to the entry in the
analyzer.
If the entries differ, a mes‑
sage will be generated.
ON/
OFF
GROUP
Select ON or OFF using
keys.
Press
enter
to confirm.
Function as described
above
ON/
OFF
Sensor Verification (TAG, GROUP)
When Memosens sensors are calibrated in the lab, it is often useful and sometimes
even mandatory that these sensors will be operated again at the same measuring
points or at a defined group of measuring points. To ensure this, you can save the
respective measuring point (TAG) or group of measuring points (GROUP) in the
sensor. TAG and GROUP can be specified by the calibration tool or automatically
entered by the transmitter. When connecting an MS sensor to the transmitter, it can
be checked if the sensor contains the correct TAG or belongs to the correct GROUP.
If not, a message will be generated, Sensoface gets “sad”, and the display backlight‑
ing turns magenta (purple). The “sad” Sensoface icon can also be signaled by a
22 mA error current. Sensor verification can be switched on in the Configuration in
two steps as TAG and GROUP if required.
When no measuring point or group of measuring points is saved in the sensor, e.g.,
when using a new sensor, Stratos enters its own TAG and GROUP. When sensor veri‑
fication is switched off, Stratos always enters its own measuring point and group.
A possibly existing TAG/GROUP will be overwritten.