Calibration by Entry of Cell Constant
You can directly enter the value for the cell constant of a sensor. This
value must be known, e.g. determined beforehand in the laboratory.
The selected process variable and the temperature are displayed.
Display
Action
Remark
Select Calibration.
Press
enter
to proceed.
Select CAL_CELL
calibration method.
Press
enter
to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Enter cell constant.
Press
enter
to proceed.
The selected pro-
cess variable and
the temperature are
displayed.
The device shows the
calculated cell constant
(at +25 °C/+77 °F).
Sensoface is active.
Use the arrow keys to
select:
• MEAS (end)
• REPEAT
Press
enter
to proceed.
End:
HOLD is deactivated
after a short time.