93
Display
Action
Remark
Display of selected
process variable
(here: mS/cm).
Now the device is in
HOLD mode:
Reinstall the sensor
and check whether the
measurement is OK.
MEAS ends calibration,
REPEAT permits
repetition.
With MEAS selected:
End calibration by
pressing
enter
.
Display of conductiv-
ity and temperature,
Sensoface is active.
After end of calibra-
tion, the outputs
remain in HOLD
mode for a short
time.
After display of
GOOD BYE, the
device automatically
returns to measuring
mode.
Note:
• Be sure to use known calibration solutions and the respective
temperature-corrected conductivity values (see table on calibration
solution).
• During the calibration procedure the temperature must be kept
constant.
Calibration with Calibration Solution